AUTO ZERO TRACKING WINDOW (AZSM) - PARAMETER 23
Description: Adjusts the percentage of a graduation that can be tracked
off during Auto Zero Tracking.
ACCEPTED PARAMETER VALUES AUTO ZERO
1-99 (default: 60) Track off 1 to 99% of a graduation (d)
100 Track off 1 graduation (1d)
200 Track off 2 graduation (2d)
300 Track off 3 graduations (3d)
POWER ON ZERO (IZSM) - PARAMETER 45
Description: Automatically zeroes the scale on power up. The maximum
range is set to +/- 10% of Scale Capacity.
If the initial load on the scale exceeds +/- 10% of Scale
Capacity, no initial scale zero will take place.
PARAMETER VALUE POWER UP ZERO
0 (default) OFF
1 ON
DISABLE ZERO KEY - PARAMETER 111
Description: Prevents the user from zeroing the scale. The [ZERO] key
will continue to perform its other functions (In Calibration
Mode, etc.).
PARAMETER VALUE ZERO KEY
0 (default) Enabled.
1 Disabled.
Auto Zero Tracking (Parameters 22 & 23) is subject to local
regulatory requirements.