HIGH SPEED DIGITAL WEIGHT INDICATORS
46
5P
eeeeee
8P
1
MORE SET-UP PARAMETERS
POWER ON UNITS - PARAMETER 6
Description: Selects the weighing units the M2000 will default to when powered
on.
Example: For the M2000 to start up weighing in pounds, set this parameter
value to “0”.
PARAMETER VALUE POWER ON UNITS
0 lbs
1 (default) kgs
SCALE OVERLOAD SETTING - PARAMETER 8
Description: Assigns the number of graduations for Scale Over. At this point
beyond True Scale Capacity, the M2000 displays EEEEEE.
Default Value: 1 Graduation. (See Parameter 3 – Page 30)
EXAMPLE: SCALE CAPACITY: 1000 lbs
GRADUATION SIZE: 2 lbs
DESIRED SCALE OVERLOAD THRESHOLD: 1010 lbs
1. Enter Parameter 8.
2. Enter 5 for the parameter value. 5 Grads at 2 lbs per Grad equals 10
lbs (5 x 2lbs).
3. At 10 lbs over scale capacity (1010 lbs) the M2000 will display its
Overload message on the screen.