EasyManuals Logo

Abbott i-STAT Alinity User Manual

Abbott i-STAT Alinity
296 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #113 background imageLoading...
Page #113 background image
7. Immediately close the cartridge and insert it into an instrument – it is important to prevent exposing
the soluon to room air since this will alter the results.
Transferring the control soluon: syringe with blunt p needle
Plain syringes (1cc or 3cc sterile syringes with blunt p needles) are recommended to transfer aqueous
control soluons from the ampule to the cartridge.
To use a syringe:
1. Place the end of the blunt p needle into the boom of the ampule.
2. Slowly remove approximately 1cc of soluon from the ampule into the syringe.
If air is trapped between the leading edge of the soluon and the plunger, do not invert the
syringe to expel it; this will not aect soluon near the p of the syringe.
If air bubbles are connually drawn into the syringe, or if a bubble is trapped near the p of the
syringe, discard the ampule and syringe.
3. Remove syringe from the ampule and expel one or two drops from the syringe.
4. Place needle into the sample cartridge and ll the cartridge to the ll mark.
5. Close cartridge and begin tesng immediately.
Transferring the control soluon: capillary tube
Plain capillary tubes are recommended to transfer aqueous control soluon from the ampule to the
cartridge. Capillary tubes with sucient ll capacity are required. For example, when lling a cartridge
that requires 95 µl, use a 150 µl capillary tube.
To use a capillary tube:
1. Place a clean, dry nger over one end of the tube and insert open end of the tube into the boom of
the ampule.
2. When the open end of the tube touches the boom of the ampule, p the ampule slightly. Remove
nger from the other end to allow lling by capillary acon. Completely ll the tube with the control
soluon.
3. Place a nger over the open end of the tube. Remove the tube from the ampule.
4. Place open end of the tube into the sample cartridge. Remove nger to allow the control soluon to
ll the cartridge to the ll mark.
5. Close cartridge immediately aer lling.
6. Tesng should be started immediately.
Reacve ingredients for i-STAT Control materials:
Analyte Control Level 1 Control Level 2 Control Level 3
Na (mmol/L) 127 141 169
K (mmol/L) 3.1 4.0 6.8
Cl (mmol/L) 85 100 122
Glu (mmol/L) 2.5 7.3 17
Urea (mmol/L) 18 4 2.7
iCa (mmol/L) 1.6 1.3 0.8
75
i-STAT Alinity — System Operaons Manual Art: 745533-01 Rev. H Rev. Date: 02-Nov-2022

Table of Contents

Other manuals for Abbott i-STAT Alinity

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Abbott i-STAT Alinity and is the answer not in the manual?

Abbott i-STAT Alinity Specifications

General IconGeneral
BrandAbbott
Modeli-STAT Alinity
CategoryMedical Equipment
LanguageEnglish

Related product manuals