EasyManua.ls Logo

Abbott i-STAT Alinity - Page 113

Abbott i-STAT Alinity
296 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
7. Immediately close the cartridge and insert it into an instrument – it is important to prevent exposing
the soluon to room air since this will alter the results.
Transferring the control soluon: syringe with blunt p needle
Plain syringes (1cc or 3cc sterile syringes with blunt p needles) are recommended to transfer aqueous
control soluons from the ampule to the cartridge.
To use a syringe:
1. Place the end of the blunt p needle into the boom of the ampule.
2. Slowly remove approximately 1cc of soluon from the ampule into the syringe.
If air is trapped between the leading edge of the soluon and the plunger, do not invert the
syringe to expel it; this will not aect soluon near the p of the syringe.
If air bubbles are connually drawn into the syringe, or if a bubble is trapped near the p of the
syringe, discard the ampule and syringe.
3. Remove syringe from the ampule and expel one or two drops from the syringe.
4. Place needle into the sample cartridge and ll the cartridge to the ll mark.
5. Close cartridge and begin tesng immediately.
Transferring the control soluon: capillary tube
Plain capillary tubes are recommended to transfer aqueous control soluon from the ampule to the
cartridge. Capillary tubes with sucient ll capacity are required. For example, when lling a cartridge
that requires 95 µl, use a 150 µl capillary tube.
To use a capillary tube:
1. Place a clean, dry nger over one end of the tube and insert open end of the tube into the boom of
the ampule.
2. When the open end of the tube touches the boom of the ampule, p the ampule slightly. Remove
nger from the other end to allow lling by capillary acon. Completely ll the tube with the control
soluon.
3. Place a nger over the open end of the tube. Remove the tube from the ampule.
4. Place open end of the tube into the sample cartridge. Remove nger to allow the control soluon to
ll the cartridge to the ll mark.
5. Close cartridge immediately aer lling.
6. Tesng should be started immediately.
Reacve ingredients for i-STAT Control materials:
Analyte Control Level 1 Control Level 2 Control Level 3
Na (mmol/L) 127 141 169
K (mmol/L) 3.1 4.0 6.8
Cl (mmol/L) 85 100 122
Glu (mmol/L) 2.5 7.3 17
Urea (mmol/L) 18 4 2.7
iCa (mmol/L) 1.6 1.3 0.8
75
i-STAT Alinity — System Operaons Manual Art: 745533-01 Rev. H Rev. Date: 02-Nov-2022

Table of Contents

Other manuals for Abbott i-STAT Alinity

Related product manuals