Cal Ver Test Out of Range Sengs (Manual Pass/Fail Determinaon Only)
The sengs in this secon determine the operator acon when a
Cal Ver result falls outside of the acceptable range and Pass/Fail
Determinaon is set to Manual.
Comment Code
Enable Select this check box to prompt the operator to enter a comment
code when results are out of range. If you select this check box to
enable prompng, choose one of the following opons:
Oponal Operator is prompted to enter an
oponal comment code.
Mandatory Operator is prompted to enter a
mandatory comment code.
Comment Style Determines whether comments must be chosen from a selecon list
or can be entered as text. Choose either of these opons:
Comment List Comments can only be chosen from the
selecon list.
Comment List with text
box
Operator will be prompted to choose a
comment or to type a comment into
the text box provided. It is a one or the
other selecon by the operator. A
maximum of 6 comments can be
displayed with a text box.
Comments A list of comments created by the CWi
user. Operator selects the most
appropriate. The comment will be
aached to the test record. Each
comment can be 20 characters in length
including spaces.
eVAS Type The le type uploaded onto the iSTAT Alinity for the use of the
Pass/Fail Determinaon feature. Select either:
eVAS Select for all countries except Germany.
ReVAS Select only if you are in Germany.
196
i-STAT Alinity — System Operaons Manual Art: 746300-01 Rev. N Rev. Date: 02-Nov-2022