Entry, mandatory
Prompt the operator for a mandatory entry. Cartridge tesng
cannot connue unl informaon is entered.
Selecon List
Click Edit Selecon List. This displays the next part of the screen:
Selecon Lis
t Entry Enter the name of the venlator mode. An example of a venlator
mode is AC. If no STATNotes items will be assigned to the page
select: No Page. If STATNotes Items are to be assigned to the page
select: STATNotes Page.
STATNotes Items In the box directly below STATNotes Items click the arrow to display
the drop down list. Select all items that should be used on the
instrument to prompt for responses for the venlator mode being
built. Items must be unique and chosen only once.
Repeat this process for all STATNotes items that will be needed for the v
enlator mode that is being
built. Once all STATNotes items have been selected for the venlator mode click Save.