Format Determine the format used to enter data. Select either:
Numeric Only Data will be entered only as numbers.
Alphanumeric Data will be entered as either leers or
numbers.
Entry Type Determine how the data will be entered into the i‑STAT Alinity instrument. Select
one of the following:
Allow Scan and Manual Entry Allow barcode scan or manual data entry.
Scan Only Allow barcode scan data entry only.
Manual Entry Only Allow manual data entry only.
Minimum Length Enter the minimum data characters
allowed.
Maximum Length Enter the maximum data characters
allowed.
Scan Mask Barcode scanning seng that allows for the selecon of specic character posions
to be retained.
• The rst posion (the posion farthest to the le) is numbered posion "1".
• The scanmask eld will accept a text input consisng of numbers 1-9, commas
(','), and hyphens ('-') to idenfy the scanmask selecon.
• If, for example, the barcode reads "abcdefghi1234567890" and the selecon
paern is "2,5-8,11-14" then the barcode will be interpreted as "befgh2345". If
the selecon paern is "3-4", then the barcode will be "cd".
• Individual posion selecons that are separated by a comma such as
"2,4,7,8,19" cannot be duplicated.
• In each number range, the starng number must be smaller than the ending
number. For example, "3-5" is a valid range, but "5-3" is not.
• Selecon ranges cannot overlap. The end of each number range must be smaller
than the beginning of the next number range. For example, "3-6,7-9" is a valid
selecon, but "3-6,6-9" is not.
• Entering consecuve commas or hyphens is invalid. For example, "2--5",
"2-5,,7-10", and "2,-5" are all invalid ranges.
232
i-STAT Alinity — System Operaons Manual Art: 746300-01 Rev. N Rev. Date: 02-Nov-2022