View Acon Ranges and Analyte Info
An individual buon is displayed for each analyte available for
the i‑STAT Alinity system. Those buons greyed out require
acvaon via a customizaon prole or are not available at
this me. Shown with the analyte is the unit of measure. Units
of measure maybe modied via a customizaon prole or on
the instrument via More Opons > Instrument Opons >
Instrument Sengs > Set Units
Touching an acve analyte buon displays a second page of
informaon:
Measurement range - Default is displayed unless a new range
is set via a customizaon prole.
Crical Test - Default of no will be displayed unless modied
via a customizaon prole.
Touching the View Reference Range buon will reveal the
ranges set for this analyte via a customizaon prole.
Note: No ranges are the default.
Touching View Acon (Crical) Ranges will reveal the ranges
set for this analyte via a customizaon prole.
Note: No ranges are the default.
Instrument Status:
Important informaon such as rmware version, barometric
pressure and prole name are listed here.
Quality Opons: Quality Control
Perform
Unscheduled QC
When liquid QC tesng is not scheduled via a customizaon
prole or if an addional liquid QC tesng event is desired,
this is the tesng pathway. Follow the prompts on the
screen to perform tesng.
Scheduled QC If liquid QC tesng has been set via a customizaon prole,
this is the tesng pathway. Follow the prompts on the
screen to perform tesng.
Perform Cartridge
QC
This feature allows liquid control tesng to release
cartridges, based on lot number, for paent or prociency
tesng. This method of QC tesng must be set up via a
customizaon prole as well as acvaon in the data
management system.
14