122 1100 Series FD Reference Manual
4 Troubleshooting and Test Functions
Figure 49 RAMAN value
Figure 50 Raman Test - Run Time Checklist (part of method)
macro flramtst.mac; seeIfReportLoaded;initNoiseDetermination
This line loads the raman macro, starts loading the noise evaluation
generator and defines ASTM noise performance parameters based on
ASTM interval definition of 0.5 minutes.
macro "report.mcx";LEO_FLD_evaluate_Noise_Test
This line starts the FLD ASTM calculation and restores original ASTM
noise calculation based on 1 minute intervals