EasyManua.ls Logo

Agilent Technologies 1100 Series - Exchanging a Flow Cell

Agilent Technologies 1100 Series
346 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Troubleshooting and Test Functions 4
1100 Series FD Reference Manual 123
Interpretation of the Results
If the test shows low Raman values, check for:
correctly positioned flow cell,
clean flow cell (flush),
no air bubble(s) (check via fluorescence scan or visual check of
cell/cuvette),
solvent inlet filter (may create air bubbles in flow cell),
clean water.
Figure 51 Raman Test - FLD Settings (part of method)
Set the
Stoptime = 23
minutes, see
on page 121.

Table of Contents

Other manuals for Agilent Technologies 1100 Series

Related product manuals