182 1100 Series FD Reference Manual
5 Repairing the Fluorescence Detector
8 Replace the FLF board and fix the screws (A and B) and
connectors (C and D).
9 Check that the safety switch is positioned correctly. Look
through board and casting.
10 Reconnect the wires to the connector. Use a flat screw
driver to release the wire socket.
11 Put the wires into the released contacts using the
tweezers.
Order of wires (left to right):
Next Steps:
• Reinstall the optical unit as described in “Installing the Optical Unit" on page 194.
• If you replaced a G1321-66501 FLF board by a G1321-66511 FLF board, update the FLD firmware to revision A.03.7x or
above (to make it compatible).
A
B
C
D
green
empty
brown
yellow
white