Lesson 1 Analog Modules
122
Local DSP
Each measure unit has DSP hardware (called local DSP), which
can be used for processing data captured by the digitizer of the
measure unit, instead of the processing by the system controller.
As the local DSP processes the captured data and judges the pass/
fail results, uploading bulky capture data to the system controller
is not needed, and thus, the upload time can be reduced. The local
DSPs in the measure units can simultaneously process the
respective captured data and judge the respective pass/fail results.
The local DSPs can minimize the test time especially in multi-site
testing, where multiple measure units operate simultaneously. In
the default setting, the local DSPs are not used.
The DSP memory is used by the local DSP during the processing,
and is used to store DSP programs that are automatically
generated from your test method programs and standard
testfunctions.
(The source unit does not have DSP hardware.)
Signal Routing in Measure Unit
For per MESA and for four
MESAs in Source&Measure
Module
The following table shows the signal routing per measure unit, and
the signal routing in the four measure units of the multi-site
baseband analog source & measure.
Function Pins in MCA Source&Measure Module
Audio DGT
Single-end-
ed input
Per measure unit:
Max. 2 inputs simultaneously of 4 inputs: (X+ or X-) and (XX+ or XX-)
Per Source&Measure module (per four measure units):
Max. 8 inputs simultaneously of 16 inputs:
(Unit5 E+ or E-), (Unit6 F+ or F-), (Unit7 G+ or G-), (Unit8 H+ or H-),
(Unit5 EE+ or EE-), (Unit6 FF+ or FF-), (Unit7 GG+ or GG-), and (Unit8 HH+ or HH-)
Audio DGT
Differential
input
Per measure unit:
Max. 2 inputs simultaneously of 2 inputs: (X+&X- pair) and (XX+&XX- pair)
Per Source&Measure module (per four measure units):
Max. 8 inputs simultaneously of 8 inputs:
(Unit5 E+&E- pair), (Unit6 F+&F- pair), (Unit7 G+&G- pair), (Unit8 H+&H- pair),
(Unit5 EE+&EE- pair), (Unit6 FF+&FF- pair), (Unit7 GG+&GG- pair), and (Unit8 HH+&HH- pair)
Video DGT
normal
mode
Single-end-
ed input
Per measure unit:
1 input at a time of 2 inputs: X+ or X-
Per Source&Measure module (per four measure units):
Max. 4 inputs simultaneously of 8 inputs:
(Unit5 E+ or E-), (Unit6 F+ or F-), (Unit7 G+ or G-), and (Unit8 H+ or H-)