EasyManua.ls Logo

ANATEL A643a - Fig 7-6: Alarm Setup Menu; Fig 7-7: Change Alarm Limit; Fig 7-8: TOC Alarm Limit Screen

ANATEL A643a
222 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Anatel A643a - Anatel A643a Alarms 103 of 220
RPS - June 2007 - Edition 12
Anatel Operator Manual
Note:
A Cond Alarm only appears if Uncompensated and Conductv both are checked in the Display
Units Menu (see “Display Units” on page 56).
7) Use the Up and Down Keys to choose the desired alarm type, indicated by an arrow.
8) Press Enter to toggle the Cond Alarm selection on (preceded by a checkmark) and off
in order to enable and disable this type of alarm (see “Uncompensated Conductivity
Alarms” on page 106). Enabling the TOC Alarm type causes the corresponding TOC
Alarm Limit to be displayed.
9) Use the Up and Down Keys to specify which digit is to be changed.
10) Press Enter to enable the Controllers Edit Mode and the flashing block cursor
becomes an underscore.
1 SENSOR NAME
X TOC Alarm:
Cond Alarm
Fig 7-6 : Alarm Setup Menu
TOC Alarm Triggers an alarm indication when the Analyzer’s reported TOC reading
exceeds its specified limit.
Cond Alarm Triggers an alarm when the instrument’s uncompensated conductivity
exceeds the USP limit.
1 SENSOR NAME
TOC Alarm Limit
05
00 ppb
S/T to change
Fig 7-7 : Change Alarm Limit
Fig 7-8 : TOC Alarm Limit Screen

Table of Contents

Related product manuals