11) Again use the Up and Down Keys to scroll until the desired value is displayed.
12) Press Enter to retain the pair’s setting.
13) Repeat the editing process with the other pair of digits, if necessary, to specify the
Analyzer’s TOC Alarm Limit (0000 to 1000 ppb). Setting the Limit to 0000 disables
the alarm.
14) Press Esc repeatedly to exit this function and return to the normal display.
7.2.2 Beeper Setup
Default: Enabled
The Beeper Setup is a global parameter for sounding the Controller’s audible alarm in the
event of a TOC limit excursion by one of its connected instruments. The alarm continues to
sound until acknowledged by pressing the Alarm Key.
To enable/disable the Controller’s Beeper:
1) With any Channel View displayed, press the Setup Key and its menu selections are
presented.
2) Use the Up and Down Keys to specify C80 Setup.
1 SENSOR NAME
TOC Alarm Limit
0250 ppb
S/T to change
Fig 7-9 : Alarm Limit Changed
Fig 7-10 : Beeper Setup
Contrast Adj:
X Beeper Setup:
Fig 7-11 : Beeper Setup Menu