Anatel A643a - Anatel A643a Setup 79 of 220
RPS - June 2007 - Edition 12
Anatel Operator Manual
6) Press Enter to initiate a single analysis. Sampl reflects the Sample Time interval
specified in the Analyzer’s setup (see “Sample Time” on page 63). Note this if the
sample interval is to be matched. The instrument’s analysis State and values are
reported as soon as they become available.
7) Press Esc to return to the Online Samples Menu.
8) Use the Up and Down Keys to specify Manual.
9) Press Enter to access its parameters and initiate a single analysis. The Duration of the
sampling period is reported during the course of the fill.
10) Press Enter to stop sampling and the analysis continues with the Oxidation
(measurement) stage. Once oxidation is complete, the sample’s TOC value is
displayed.
The Manual Mode can be terminated at any time by pressing the Esc Key to exit the screen.
1 SENSOR NAME
State SAMPLE
Sampl: 0:01:00
X
Manual:
Fig 5-51 : Sample Time Interval
Fig 5-52 : Manual Oxidize State