86 of 220 Calibration and Validation - Anatel A643a
RPS - June 2007 - Edition 12
Operator Manual Anatel
17) Press Enter to accept the new calibration and initiate a printout of the data. If the
percent change in the slope exceeds ±15% of factory calibration, the calibration is
deemed “unacceptable.” A opportunity to repeat the calibration procedure using new
sucrose standards will be given.
18) Once the calibration procedure is complete, remove the standard vial and reinstall the
sample vessel before proceeding.
A printout or display of the calibration information also is available through the Data History
Menu selection (see “Data History” on page 99).
Fig 6-7 : TOC Calibration Results
Fig 6-8 : Sample TOC Calibration Printout