Anatel A643a - Analog Outputs 127 of 220
RPS - June 2007 - Edition 12
Anatel Operator Manual
6) Again use the Up and Down Keys to scroll the pair until the desired value is displayed.
7) Press the Esc Key to retain the setting.
8) Repeat the editing process with the other pair of digits if necessary to display the
desired value.
9) Press Esc twice to retain the Zero-Scale TOC setting and return to the DAC Range
Menu.
10) Select DAC Full-Scale and press Enter.
Note:
The Full-Scale TOC value must be greater than the Zero-Scale TOC value.
11) Specify TOC DAC once again and then press E to display its full-scale parameter
screen. The flashing block cursor highlights the first pair of digits.
12) Use the same procedure as that described in step 4 through step 9 to specify the Full-
Scale TOC value and return to the DAC Range Menu.
1 SENSOR NAME
Zero-Scale TOC:
0025 ppb
S/T to change
Fig 9-12 : Zero-Scale TOC Data
Fig 9-13 : Full-Scale TOC Screen
1 SENSOR NAME
Full-Scale TOC:
05
00 ppb
S/T to change
Fig 9-14 : Full-Scale TOC Data