Anatel A643a - Anatel A643a Maintenance 165 of 220
RPS - June 2007 - Edition 12
Anatel Operator Manual
To reset the instrument’s UV Lamp counter:
20) Select the appropriate Analyzer in any View and press Setup to access its submenu.
21) Specify Sensor Setup and press Enter.
22) Specify Analysis Setup and press Enter.
23) Specify More Setup and press Enter.
24) Specify Cell Setup and press Enter.
25) Specify Lamp Install and press Enter.
26) Press Enter to confirm the new Lamp installation.
27) Press Esc repeatedly to exit the screen and return to the main menu.
Fig 13-13 : Lamp Setup
Fig 13-14 : Resetting the Lamp Counter