EasyManuals Logo

ANATEL A643a User Manual

ANATEL A643a
222 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #96 background imageLoading...
Page #96 background image
96 of 220 Calibration and Validation - Anatel A643a
RPS - June 2007 - Edition 12
Operator Manual Anatel
7) Use the Up and Down Keys to scroll until the sucrose standard value found on the
Certificate of Analysis is displayed.
8) Press Enter to retain the setting and the cursor advances to the 1,4-benzoquinone
standard.
9) Again use the Up and Down Keys to enter the certified 1,4-benzoquinone standard
value from the Certificate of Analysis, then press Enter.
10) Press Enter to confirm both Certificate of Analysis values have been properly
entered.
11) Remove the empty sample vessel and insert the blank vial. Press Enter and its
contents are sampled and analyzed. The results of each of the analysis replicates are
reported.
CAUTION:
Some instrument surfaces may be hot if the sample water is hot. Allow surfaces to cool before
touching.
CAUTION:
Use extreme care when installing and removing the sample vessel and vials. Contact with the
exposed needle could cause serious injury.
Fig 6-25 : Certificate of Analysis Values
Enter “C of A"
values
508 ppb sucrose
500 ppb benz
Fig 6-26 : Certificate of Analysis Displayed

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the ANATEL A643a and is the answer not in the manual?

ANATEL A643a Specifications

General IconGeneral
BrandANATEL
ModelA643a
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals