Chapter 6: Shutdown and Maintenance 185
Prime After Tank Refill
Use the Prime After Tank Refill command to prime the fluid lines if a fluidics
container was disconnected for refilling.
1 Open the flow cell access door and turn off the stream.
2 Select Cytometer > Cleaning Modes > Prime After Tank Refill.
3 Select the checkboxes for the tanks that were refilled, then click OK.
The cytometer proceeds with priming the specified tanks. The bleach and
ethanol tanks are primed before the DI tank is primed. A progress message
appears while the tanks are being primed.
4 Click OK when the tank prime is complete.