Chapter 6: Shutdown and Maintenance 195
4 Detach the sheath tubing and the ethanol or bleach tubing from their
respective ports on the fluidics cart.
Figure 6-3 Sheath and ethanol ports
5 Plug the ethanol or bleach tank quick-disconnect into the sheath port.
Cleaning the Lines with Ethanol or Bleach
1 Install a tube containing ethanol or bleach onto the loading port.
2 Select Cytometer > Cleaning Modes > Prepare for Aseptic Sort.
A dialog appears with a reminder of the steps required to set up for
cleaning. Confirm that these steps have been done.
bypass tubing
sheath tubing
ethanol tubing