Chapter 6: Shutdown and Maintenance 215
Figure 6-10 Filter outlet tubing connected to sheath line
3 Connect the sheath line to the filter purge tubing (Figure 6-11).
Figure 6-11 Filter purge tubing connected to sheath line
4 Turn on the stream.
5 Gently tap the bubble filter to ensure that there are no large air bubbles left
in the filter.
6 Turn off the stream.
sheath line
filter outlet tubing
quick-disconnect coupling
filter purge tubing
sheath line

Table of Contents