22.6 Specifications
Analyzer with Autosampler
Analyzer with Autosampler
Closed and open vial mode: 110 l
Human whole blood (K
3
-EDTA anticoagulant)
By means of the front panel keyboard (enter ID)
By means of the barcode labels (manual and/or auto-sampler)
Ceramic shear valve with 3 separate primary loops
CBC+5DIFF mode (24 parameters): WBC, LYM, MON, NEU, EOS, BAS, LYM%,
MON%, NEU%, EOS%,BAS%,RBC, HCT, MCV, RDWsd, RDWcv HGB, MCH, MCHC,
PLT, MPV, PDWsd, PDWcv, PCT
Volumetric impedance change for WBC, RBC, PLT
Light scattering BASO measurement
Light scattering 4-diff measurement: LYM, MON, NEU, EOS
Spectrophotometry for HGB
WBC: 80 m, RBC/PLT: 70 m
WBC: 80 m, RBC/PLT: 70 m
Light source: green LED with 568 nm wavelength
Detector: light to frequency converter
Light source: semiconductor laser diode with 650 nm wavelength and 7mW
Beam divergence: ~1.5°
Class 1 Laser Product (Class IIIB laser module if the protective housing is
opened, and interlocks defeated)
IEC 60825-1: Ed. 2(2007)
Quartz flow cell with hydro-dynamic focusing
Detector: fiber optic coupled PIN Si photodiodes
Internal safety interlock
Horizontal calibration of laser beam position
Fine calibration with 5 m polystyrene microsphere calibration material
Diatro•Dil-5P (20 liter)
Diatro•Diff-5P (1 liter)
Diatro•Lyse-5P (5 liter)
Diatro•Hypocleaner CC (100 ml) (Emergency cleaner)
WBC/BAS dilution - 1:170
RBC/PLT dilution - 1:21,250
4 Diff dilution - 1:50
Diacon 5, Manufacturer: Diatron