154
Parameter
AMR
Units
Low
High
MCHC
NA
NA
g/dl
RDWcv
8
40
%
RDWsd
10
250
fl
PLT
16
1000
10
3
cells/µl
MPV
4.0
30.0
fl
Table 34. Analytical Measuring Range
22.7.10 Interfering Substances
The following substances can interfere with parameter measurement and alternate measurement procedures
may be required.
Parameter
Interference
WBC
> 5 NRBCs/100 WBCs , PLT clumps/large PLTs
RBC
WBC Count > 75.0 x10
3
/µL
MCV
WBC Count > 75.0 x10
3
/µL
PLT
PLT clumps/largePLTs
Hemoglobin
WBC count > 75.0 x10
3
/µL , Lipids > 280 mg/dL
Differential
> 5 NRBC’s/100 WBC’s , PLT clumps/largePLTs