To prepare the analyzer for shipping, storage or extended periods of inactivity, please drain
the reagents and repackage the ‘Abacus 5’ in its original packaging. Do not expose the
‘Abacus 5’ to direct sunlight, extreme temperature or humidity (>80%).
The analyzer operates with chemically and biologically active reagents. Physical contact
with these reagents should be avoided. Please read reagent descriptions carefully for
possible emergency actions.
To ensure reliable operation and reliable results:
Only human blood samples should be analyzed
Only genuine Diatron reagents should be used
Required maintenance (user and service level) should be performed as
recommended in this manual
Only Diatron certified service personnel should perform service actions
Only genuine Diatron service materials and spare parts should be used
Genuine reagents and service materials and spare parts are available from Diatron.
Only Diatron certified service personnel that have successfully completed the ‘Diatron
Abacus 5 Service Training’ program are qualified to service the ‘Abacus 5’ analyzer.
Before operating the ‘Abacus 5’ analyzer, all operators should complete a ‘Diatron Abacus
5 Operator Training’ program. This program is offered by Diatron or by Diatron certified
service personnel.
Replacement materials or spare parts (tubes, valves, etc.) which might have been in
contact with human blood or reagents should be handled as a potentially biologically
hazardous and chemically dangerous material. All applicable laws and regulations must be
observed in the handling and disposal of these materials.
The ‘Abacus 5’ is designed for laboratory operation. Mobile operation is not supported.
Operate ‘Abacus 5’ within the ambient temperature range described in section 2.4.
The IVD equipment complies with the emission and immunity requirements described in
relevant part of the IEC 61326 series.
This equipment has been designed and tested to CISPR 11 Class A. In a domestic
environment it may cause radio interference, in which case, you may need to take
measures to mitigate the interference.
Electromagnetic environment should be evaluated prior to operation of the device.
This analyzer contains electronic components. Please handle electronic waste
adhering to local or federal regulations.
CAUTION – Use of controls or adjustments or perfomance of procedures other than those
specified herein may result hazardous radiation exposure.