Note
Temperature is only available for input A.
Procedure
1.
From the Analyze main menu, press F1 Manual Analyze > F12 Input Setup > F7 Sensor
Setup.
You can also access the Sensor Setup option from the ALT2 Analyze main menu.
2. Set the following options as necessary for each input.
Option Description
Change Sensor Type Select the type of sensor to use for the input.
Change Sensitivity Enter the sensitivity in volts per engineering unit, such as g's for an
accelerometer.
Change Sensor
Power
Select ON to use the accelerometer input where the analyzer powers
the sensor. Select OFF to use the volts input, and the analyzer will not
provide power to the sensor.
Change Signal
Coupling
The coupling mode for sensor power ON is always AC coupled. When
sensor power is OFF, the mode can be changed to AC or DC.
Change Accel Config Set the accelerometer sensors as single axis (one signal), biaxial (two
signals), or triaxial (three signals).
3. Press ALT to set the parameters for input C and input D.
4.
Press Enter.
6.5.3 Set the data units in Analyze
Set the data units for the spectra and waveform for each selected input. The available
options vary based on the measurement type you selected. Based on the Analyze mode
you select, the data units may change.
Procedure
1. From the Analyze main menu, press F1 Manual Analyze > F12 Input Setup.
2. For each input, select the units.
3. Press Enter.
Analyze and Advanced Analyze
MHM-97432 Rev 7 115