EasyManuals Logo

Emerson CSI 2140 User Manual

Emerson CSI 2140
404 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #255 background imageLoading...
Page #255 background image
2. From the Review Measurements screen, press F8 Data Detail.
3.
Press F1 Display Sine Fit to display the sine fit data plot.
Sine curve data plotFigure 7-47:
On the sine curve data plot, the amplitude versus degree data is shown.
When a cursor is active, the following data values for the selected point are
displayed below the plot:
• Point—Number of the selected point.
• Angle—Angular position of the selected point.
• Vert—Actual vertical position of the selected point. The background is green for
an active point and red for a deleted point.
• Est—Amplitude value of the sine fit curve at the particular angular position.
The following calculated sine curve values are displayed at the bottom of the
analyzer screen:
• Phase—The plus zero transition position.
• Offset—The centerline of the laser beam travel on the sensor in the radial
direction. The initial position of the laser beam on the sensor affects this value.
• Amplitude—The maximum displacement of the sine curve from the offset line.
• Correlation—The confidence factor in percent that is used to determine the
quality of the sine curve data.
Advanced Laser Alignment
MHM-97432 Rev 7 249

Table of Contents

Other manuals for Emerson CSI 2140

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Emerson CSI 2140 and is the answer not in the manual?

Emerson CSI 2140 Specifications

General IconGeneral
BrandEmerson
ModelCSI 2140
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals