Option Description
F9 Set Fixed Ref
Input
Set the reference input to A, B, C, or D. Do not move the fixed reference
sensor during the job. The default is input A. If you disable the reference
input on the Input Setup screen, "???" displays next to the F9 Set Fixed Ref
Input key. For Modal jobs, the reference input is the input for the hammer.
F10 Set Fixed Ref
Point
Enter a value between 1 and the maximum number of points you entered.
This sets the measurement point of the fixed reference. The default is 1.
F11 Set Fixed Ref
Direction
Set the measurement direction (also referred to as the measurement
coordinate or measurement axis) of the fixed reference. The available
directions are dependent on the global coordinate system and directions
defined for the job. The default is X.
F12 Set Fixed Ref
Polarity
Set the direction for a measurement. The direction is usually determined
by the construction of the equipment. If you cannot place the sensor in
the positive direction, specify a negative direction. The default is Positive.
5. Press Enter.
9.6.6 Add a measurement point to an ODS/Modal job
You can add a point after you have collected data for a job.
Procedure
1.
From the ODS/Modal main menu, press F8 Job Manager.
2. Use the up and down arrow keys to select a job.
3. Press F4 Edit This Job > F6 Add Point.
4. Enter the total number of measurement points.
5. Press Enter.
9.6.7 Add a measurement direction to an ODS/Modal point
If your measurement point has fewer than three directions, you can add additional
directions. The option below appears after you collect data for a point.
Procedure
1.
From the ODS/Modal main menu, press F8 Job Manager.
2. Use the up and down arrow keys to select a job.
3. Press F4 Edit This Job > F5 Add Meas Direction.
4. Use the up and down arrow keys to select the desired directions.
5. Press Enter.
Make sure you have your inputs set up properly. You may see "???" if an input needs
to be set up for that direction.
ODS/Modal
MHM-97432 Rev 7 291