Option Description
Send to PC Send files between the analyzer and AMS Machinery Manager. Files
can include routes, jobs, screen captures, splash screens, or summary
reports. The default is Send to PC.
Store as BMP on CARD Save the file as a .bmp to the SD memory card.
Store as JPG on CARD Save the file as a .jpg to the SD memory card.
Print Off Disable printing. Print options, such as the F7 Connect For Printing
option on the Home ALT2 screen, are not available.
3. Press Enter.
2.14.5 Set the low-battery warning level
When the battery pack discharges to the specified percent, the analyzer displays a
message telling you to recharge the battery pack.
Procedure
1.
Press Home > ALT > F2 General Setup > ALT > F3 Set Warning Level.
2. Enter a percent between 0 and 99.
A 0 disables this feature. The default is 15 percent.
3. Press Enter.
2.14.6 Configure the power key
You can set the number of seconds you must hold the power key to shut down the
analyzer. This option does not change the hold time to put the analyzer in standby.
Increase the number of seconds to prevent users from accidentally pressing the power key
and shutting down the analyzer.
Procedure
1.
Press Home > ALT > F2 General Setup > ALT > F4 Set Hold Time.
2. Enter a number between 1 and 5 seconds.
The default is 3 seconds.
3. Press Enter.
2.14.7 Set the time and date
The analyzer's time must be within 15 minutes of the computer's time for the analyzer and
computer to communicate successfully. If there is a time conflict, the analyzer prompts
you to automatically synchronize the analyzer with the computer's time.