2. Place the Balancing Compass on an axial face with the indicator needle pointing at 0
degrees.
Attach and zeroFigure 10-9:
3. Turn the rotor until the indicator reads the desired angular position from the
applicable scale. This position is now aligned with the stationary reference.
Turn the rotorFigure 10-10:
Determine the position of a sensor or tachometer pick up
from top dead center (TDC)
1.
Align a reference mark on the rotor to be balanced with the sensor, or place a
temporary reference mark on the shaft adjacent to the sensor.
Align a reference markFigure 10-11:
Balance
362 MHM-97432 Rev 7