10 Complete the configuration
10.1 Test or tune the system using sensor simulation
Display MenuService ToolsSensor SimulationSimulate Sensor
ProLink III Device ToolsDiagnosticsTestingSensor Simulation
Field communicator DiagnosticsSimulationSimulate Sensor
Use sensor simulation to test the system's response to a variety of process conditions, including boundary
conditions, problem conditions, or alert conditions, or to tune the loop.
Restriction
Sensor simulation is available only on flow meters with the enhanced core processor.
Prerequisites
Before enabling sensor simulation, ensure that your process can tolerate the effects of the simulated process
values.
Procedure
1. Enable sensor simulation.
2. For mass flow, set Wave Form as desired and enter the required values.
Option
Required values
Fixed Fixed Value
Sawtooth Period
Minimum
Maximum
Sine Period
Minimum
Maximum
3. For density, set Wave Form as desired and enter the required values.
Option
Required values
Fixed Fixed Value
Sawtooth Period
Minimum
Maximum
Sine Period
Minimum
Maximum
Configuration and Use Manual Complete the configuration
MS-00809-0200-1600 May2024
Configuration and Use Manual 121