16.11.22 [102] Drive Overrange
Alert
Drive Over-Range
Cause
The drive power (current/ voltage) is at its maximum.
Recommended actions
1. Verify that the tubes are full of process fluid.
2. Check flow, density and temperature against the values reported by the device.
3. For liquid applications, check for partially filled tubes. For gas applications, check for condensate,
compressor oil, etc.
4. Check for foreign material in the tubes, coating on tube walls, or other process problems.
5. Check the drive gain and the pickoff voltage.
Refer to Check the pickoff voltage.
6. Check sensor coil resistance readings. If you find problems, replace the sensor.
Refer to Check for internal electrical problems.
7. Ensure that the sensor orientation is appropriate for your application. Check your sensor
installation manual for proper orientation.
Settling from a two-phase or three-phase fluid can cause this alert even if the flow tubes are full.
16.11.23 [104] Calibration in Progress
Alert
Function Check in Progress
Cause
A calibration is running.
Recommended actions
Allow the test to complete.
16.11.24 [105] Two-Phase Flow
Alert
Process Aberration
Cause
Recommended actions
Check for two-phase flow.
Refer to Configure Lower Range Value (LRV) and Upper Range Value (URV) for an mA Output.
16.11.25  [110] Frequency Output 1 Saturated
Alert
Output Saturated