16.11.20 [034] Smart Meter Verification Failed
Alert
Function Check Failed and Smart Meter Verification Failed
Cause
The current Smart Meter Verification value is statistically different than the factory baseline value.
Recommended actions
1. Minimize process instability and repeat the test.
2. Contact Customer Service.
16.11.21 [035] Smart Meter Verification Aborted
Alert
Function Check Failed and Smart Meter Verification Aborted
Cause
The Smart Meter Verification test did not complete, possibly because it was manually aborted or because
process conditions were too unstable.
Recommended actions
1. Minimize process instability and repeat the test.
2. Check the Smart Meter Verification abort code and take appropriate steps.
3. Contact customer service.
Table 16-2: SMV aborts code
Abort code Abort reason Abort recommendation
1 Manual abort by end user Not available
2 SMV timeout Check status of pickoff and drive coils.
3 Pickoff Voltage low Check status of pickoff and drive coils.
4 Temperature unstable Ensure temperature is stable. Start again.
5 Cal state in progress Not available
6 Unused Not available
7 Drive loop AGC reported an amplitude error Check status of pickoff and drive coils.
8 High flow (dt) standard deviation Reduce flow rate and start again.
9 High flow (dt) mean value Reduce flow rate and start again.
10 State in progress Not available
11 Verification complete Not available
12 Transmitter in fault Clear alarms before proceeding.
13 No factory air verification Perform factory calibration on air.
14 No factory water verification Perform factory calibration on water.
15 Drive frequency drift from carrier frequency
Ensure temperature, flow, and density are
stable. Start again.
Troubleshooting Configuration and Use Manual
May2024 MS-00809-0200-1600