13.1.5 View Smart Meter Verification test results
When the Smart Meter Verification Basic test is complete, a pass/fail result is displayed. With Smart Meter
Verification Professional, detailed results and reports are available.
Note
With Smart Meter Verification Professional, the twenty most recent results are available. If viewed using
ProLink III Basic or Professional, results for all tests that are in the PC database are available.
View Smart Meter Verification test results using the display
Results of the current Smart Meter Verification Basic or Professional test display automatically after the test is
complete.
With Smart Meter Verification Professional, use the following procedure to view previous test results.
Procedure
1. Choose MenuOperationsSmart Meter VerificationRead Verification History.
2. To view Pass/Fail status time and date for an individual test, select it from the list.
View Smart Meter Verification test results using ProLink III Basic or
Professional
Results of the current Smart Meter Verification Basic or Professional test display automatically after the test is
complete.
With Smart Meter Verification Professional, use the following procedure to view previous test results.
All Smart Meter Verification test results are stored on the same PC that Smart Meter Verification was initiated
from. To view a previous test report, the Smart Meter Verification Professional test must have been run on that
same PC.
Procedure
1. Choose one of the following options:
Device ToolsDiagnosticsMeter VerificationRun TestView Previous Test Results
Smart Meter Verification OverviewMeter VerificationHistory
2. To view details, choose the results of interest: Show Report (or Next to show the report).
ProLinkIII displays a report containing details of the most recent tests. The report is automatically
saved to the Smart Meter Verification database. You can print or export the report.
View Smart Meter Verification test results using a field communicator
In addition to test results, some field communicator brands provide a trend chart.
Results of the current Smart Meter Verification Basic or Professional test display automatically after the test is
complete.
With Smart Meter Verification Professional, use the following procedure to view previous test results.
Procedure
1. Choose Device SettingsCalibrationSmart Meter VerificationManual Verification.
2. Choose Upload Results Data from Device.
The field communicator stores only the most recent test result. To view earlier results, you must upload
them from the device. They will be available only for the current session.
3. Choose Show Results Table.
The field communicator displays detailed results for the first test.