Cause
The output is set to a constant state. An output test is in progress.
Recommended actions
Check to see if an output test is in progress. If so, unfix it.
16.11.30 [120] Curve Fit Failure (Concentration)
Alert
Configuration Error
Cause
One or more errors in the concentration curve input
Recommended actions
1. Verify the configuration of the concentration measurement application.
2. Ensure the density and temperature units match those that the curve was written in.
3. Contact customer service.
16.11.31 [121] Extrapolation Alert (Concentration)
Alert
Process Aberration
Cause
The line density or line temperature is outside the range of the concentration matrix plus the configured
extrapolation limit.
Recommended actions
1. Verify the configuration of the concentration measurement application.
2. Ensure the live density and temperature readings fall within limits of the matrix being used.
16.11.32 [123] Pressure Overrange (API Referral)
Alert
Process Aberration
Cause
The external line pressure is outside the range of the API table.
Recommended actions
1. Ensure that process pressure is within the range of the API table.
2. Verify the configuration of the API Referral application and related parameters.
3. Ensure the transmitter's external pressure engineering unit matches that of the actual external
pressure device.