3.2.4 Load a configuration file using ProLinkIII
You can load a configuration file to a transmitter's internal memory, as well as loading a backup file or a
replication file. Two PC file formats are supported: the 1600 format and the ProLinkIII format.
Note
When you use ProLink III format for configuration files, you can specify configuration parameters individually
or by groups. Therefore, you can use this format for both backup and replication.
Procedure
• To load a backup file or replication file in 1600 format from the PC:
a) Choose Device Tools → Configuration Transfer → Load Configuration.
b) Select On my computer in 1600 device file format and select Next.
c) Select Restore.
d) Set the file type.
— To load a backup file, set the file type to Backup.
— To load a replication file, set the file type to Transfer.
e) Navigate to the file you want to load, and select it.
The parameters are written to the internal memory, and the new settings become effectively immediately.
• To load a file in ProLinkIII format from the PC:
a) Choose Device Tools → Configuration Transfer → Load Configuration.
b) Select On my computer in ProLink III file format and select Next.
c) Select the parameters that you want to load.
d) Select Load.
e) Set the file type to Configuration file.
f) Navigate to the file you want to load, and select it.
g) Select Start Load.
The parameters are written to the internal memory, and the new settings become effectively immediately.
3.2.5 Replicate a transmitter configuration
Replicating a transmitter configuration is a fast method to set up similar or identical measurement points.
Procedure
1. Configure a transmitter and verify its operation and performance.
2. Use any available method to save a replication file from that transmitter.
3. Use any available method to load the replication file to another transmitter.
Configuration and Use Manual Introduction to configuration and commissioning
MS-00809-0200-1600 May2024
Configuration and Use Manual 25