16.11.9 [018] EEPROM Error (Transmitter)
Alert
Electronics Failed
Cause
There is an internal memory problem with the transmitter.
Recommended actions
1. Power-cycle the transmitter to see if the alert clears.
Important
This alert will not clear until you cycle power to the meter.
2. Ensure that all wiring compartment covers are installed correctly.
3. Ensure that all wire connections are tight.
4. Ensure that all transmitter wiring meets specifications and that all cable shields are properly
terminated.
5. Ensure that all meter components are grounded properly.
6. Evaluate the environment for sources of high electromagnetic interference (EMI) and relocate the
transmitter or wiring as necessary.
7. If the alert persists, replace the transmitter.
8. Contact Customer Service.
16.11.10 [019] RAM Error (Transmitter)
Alert
Electronics Failed
Cause
There is a ROM checksum mismatch in the transmitter or the RAM address location cannot be written in the
transmitter. This alert will not clear until you reboot or power cycle the transmitter.
Recommended actions
1. Reboot or power-cycle the transmitter to see if the alert clears.
2. Ensure that all wiring compartment covers are installed correctly.
3. Ensure that all wire connections are tight.
4. Ensure that all transmitter wiring meets specifications and that all cable shields are properly
terminated.
5. Ensure that all meter components are grounded properly.
6. Evaluate the environment for sources of high electromagnetic interference (EMI) and relocate the
transmitter or wiring as necessary.
7. If the alert persists, contact Customer Service.
16.11.11 [020] Calibration Factors Missing
Alert
Configuration Error
Troubleshooting Configuration and Use Manual
May2024 MS-00809-0200-1600
194 Emerson.com