EasyManua.ls Logo

Emerson Micro Motion 1600 - 16.19.5 Perform a Frequency Output test using ProLink III; 16.19.6 Perform a Discrete Output tests using ProLink III

Emerson Micro Motion 1600
234 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
7. Read the mA current at the receiving device and compare it to the transmitter output.
The readings do not need to match exactly. If the values are slightly different, you can correct the
discrepancy by trimming the output.
8. Select UnFix mA.
Postrequisites
If the mA Output readings are within 20 microamps of each other, you can correct this discrepancy by
trimming the output.
If the discrepancy between the mA Output readings is greater than 20 microamps, or if at any step the
reading was faulty, verify the wiring between the transmitter and the remote device, and try again.
You must end the test. It will not time out by itself. A power cycle will also end the test.
16.19.5 Perform a Frequency Output test using ProLinkIII
Prerequisites
Before performing a Frequency Output test, configure Channel B as a Frequency Output.
Follow appropriate procedures to ensure that testing will not interfere with existing measurement and control
loops.
Procedure
1. Choose Device ToolsDiagnosticsTesting and select the Frequency Output test.
2. Enter the Frequency Output value in Fix to.
3. Select Fix FO.
4. Read the frequency signal at the receiving device and compare it to the transmitter output.
5. Select UnFix FO.
Postrequisites
You must end the test. It will not time out by itself. A power cycle will also end the test.
16.19.6 Perform a Discrete Output tests using ProLinkIII
Prerequisites
Before performing a Discrete Output test, configure Channel B as a Discrete Output.
Follow appropriate procedures to ensure that testing will not interfere with existing measurement and control
loops.
Procedure
1. Choose Device ToolsDiagnosticsTestingDiscrete Output Test.
2. If your transmitter is configured for two or more Discrete Outputs, determine which Discrete Output
you want to test.
3. Set Fix To: to ON.
4. Verify the signal at the receiving device.
5. Set Fix To: to OFF.
6. Verify the signal at the receiving device.
7. Select UnFix.
Postrequisites
If the Discrete Output readings are reversed, check the setting of Discrete Output Polarity.
Configuration and Use Manual Troubleshooting
MS-00809-0200-1600 May2024
Configuration and Use Manual 211

Table of Contents

Other manuals for Emerson Micro Motion 1600

Related product manuals