Refer to Check for two-phase flow (slug flow).
4. Ensure the lower puck is seated to sensor feedthrough or junction box.
5. Contact customer service.
16.11.4 [008] Density Overrange
Alert
Density Overrange
Cause
• Tubes plugged
• Incorrect density calibration parameters
• Two-phase flow
Recommended actions
1. If other alerts are present, resolve those alert conditions first.
2. Check your process conditions against the values reported by the device.
3. Check for two-phase flow by checking for two-phase alerts. If two-phase flow is the problem, alerts
will be posted.
4. Contact customer service.
16.11.5 [009] Transmitter Initializing
Alert
Transmitter Initializing
Cause
• The transmitter is in power-up mode.
• Insufficient DC startup current.
Recommended actions
1. Allow the meter to complete its power-up sequence. The alert should clear automatically.
2. Verify that the DC power supply current is sufficiently rated for the device.
16.11.6 [010] Calibration Failed
Alert
Calibration Failure
Cause
The calibration failed.
Recommended actions
1. Ensure that your calibration procedure meets the documented requirements.
2. Reboot or power-cycle the transmitter.
3. Retry the procedure. If it still fails, contact Customer Support.