Table 7-1: Default configuration for display variables
Display variable Process variable assignment
Display Variable 1 Mass flow rate
Display Variable 2 Mass total
Display Variable 3 Volume flow rate
Display Variable 4 Volume total
Display Variable 5 Density
Display Variable 6 Temperature
Display Variable 7 Drive gain
Display Variable 8 None
Display Variable 9 None
Display Variable 10 None
Display Variable 11 None
Display Variable 12 None
Display Variable 13 None
Display Variable 14 None
Display Variable 15 None
7.1.3 Configure a two-line display screen
Display
Menu → Configuration → Display Settings → Display Variables → 2-Value View
ProLink III Device Tools → Configuration → Transmitter Display → Display Variables → 2 PV Screen Slot #X
Field communicator Device Settings → Display → General Settings → Language
You can configure one display screen to show two process variables at a time. For each of these process
variables, the current value and the measurement is shown.
The two-line display screen operates like one of the basic 15 screens. You can use ˅ and ˄ to scroll to it. If Auto
Scroll is enabled, the two-line screen will be the last screen in the cycle.
7.1.4 Configure the number of decimal places (precision) shown on the
display
Display
Menu → Configuration → Display Settings → Decimals on Display
ProLink III Device Tools → Configuration → Transmitter Display → Display Variables → Decimal Places for x
Field communicator Device Settings → Display → Decimal Places
You can specify the precision (the number of decimal places) that the display uses for each display variable.
You can set the precision independently for each display variable.
The display precision does not affect the actual value of the variable, the value used in calculations, or the
value reported via outputs or digital communications.
Procedure
1. Select a process variable or a diagnostic variable.