ProLink III Device Tools → Configuration → Process Measurement → Density → Density Cutoff
Field communicator Device Settings → Process Configuration → Density → Density Cutoff
Density Cutoff specifies the lowest density value that will be reported as measured. All density values below
this cutoff will be reported as 0.
Procedure
Set Density Cutoff to the value you want to use.
• Default: 0.2 g/cm
3
• Range: 0.0 g/cm
3
to 0.5 g/cm
3
Effect of Density Cutoff on volume measurement
Density Cutoff affects liquid volume measurement. If the density value goes below Density Cutoff, the volume
flow rate is reported as 0.
4.6 Configure temperature measurement
The temperature measurement parameters control how temperature data is processed. Temperature data
is used in several different ways, including temperature compensation, API Referral, and concentration
measurement.
4.6.1 Configure Temperature Measurement Unit
Display
Menu → Configuration → Process Measurement → Temperature → Units
ProLink III Device Tools → Configuration → Process Measurement → Temperature → Temperature Unit
Field communicator Device Settings → Process Configuration → Temperature → Temperature Unit
Temperature Measurement Unit specifies the unit that will be used for temperature measurement.
Procedure
Set Temperature Measurement Unit to the option you want to use.
Default: °C (Celsius)
Options for Temperature Measurement Unit
The transmitter provides a standard set of units for Temperature Measurement Unit. Different
communications tools may use different labels for the units.
Unit description
Label
Display ProLink III Field communicator
Degrees Celsius °C °C degC
Degrees Fahrenheit °F °F degF
Degrees Rankine °R °R degR
Kelvin °K °K Kelvin
Configuration and Use Manual Configure process measurement
MS-00809-0200-1600 May2024
Configuration and Use Manual 43