2. See Figure 13-2.
Figure 13-2: D1 and D2 Calibration Work Flow with a Field Communicator
Postrequisites
If you disabled LD Optimization before the calibration procedure, re-enable it.
13.7 Adjust concentration measurement with Trim Offset
Trim Offset adjusts the meter's concentration measurement to match a reference value.
Tip
You can adjust concentration measurement by applying the trim offset only, or by applying both the trim
offset and the trim slope. For most applications, the trim offset is sufficient.