Cause
The calculated output value is outside the range of the output.
Recommended actions
1. Check the scaling of the Frequency Output.
2. Check your process conditions against the values reported by the device.
3. Ensure that both devices are using the same measurement unit.
4. Purge the sensor tubes.
16.11.26 [111] Frequency Output 1 Fixed
Alert
Output Fixed
Cause
The output is configured to send a constant value. A frequency output test is in progress.
Recommended actions
Check to see if a frequency output test is in progress. If so, unfix the output.
16.11.27 [116] Temperature Overrange (API Referral)
Alert
Process Aberration
Cause
The line temperature is outside the range of the API table.
Recommended actions
1. Ensure that process temperature is within the range of the API table.
2. Verify the configuration of the API Referral application and related parameters.
16.11.28 [117] Density Overrange (API Referral)
Alert
Process Aberration
Cause
The line density is outside the range of the API table.
Recommended actions
1. Ensure that process density is within the range of the API table.
2. Verify the configuration of the API Referral application and related parameters.
16.11.29 [118] Discrete Output 1 Fixed
Alert
Output Fixed
Troubleshooting Configuration and Use Manual