16.6.4 Unusually low density reading
Cause
Two-phase flow
Incorrect calibration factors
In low-frequency meters, erosion or corrosion
Recommended actions
1. Check for alerts, especially Drive Overrange.
2. Check your process conditions against the values reported by the device.
3. Verify all of the characterization or calibration parameters. See the sensor tag or the calibration
sheet for your meter.
4. Purge sensor tubes.
5. Check for tube erosion, especially if the process fluid is abrasive.
6. Perform Smart Meter Verification.
16.7 Discrete Output troubleshooting
16.7.1 No Discrete Output
Cause
Output not powered
Wiring problem
Circuit failure
Recommended actions
1. Check the power supply and power supply wiring if power source is set to external.
2. Verify the output wiring.
3. Verify that the channel is wired and configured as a Discrete Output.
4. Contact customer service.
16.7.2 Loop test failed
Cause
Output not powered
Power supply problem
Wiring problem
Circuit failure
Recommended actions
1. Check the power supply and power supply wiring.
2. Verify the output wiring.
3. Contact customer service.
Configuration and Use Manual Troubleshooting
MS-00809-0200-1600 May2024
Configuration and Use Manual 183