Related information
Configure mA Output Direction
Configure Frequency Output Direction
Configure Discrete Output Source
Configure totalizers and inventories
Effect of Sensor Flow Direction Arrow on digital communications
4.2 Configure mass flow measurement
The mass flow measurement parameters control how mass flow is measured and reported. The mass total
and mass inventory are derived from the mass flow data.
4.2.1 Configure Mass Flow Measurement Unit
Display
Menu → Configuration → Process Measurement → Flow Variables → Mass Flow Settings → Units
ProLink III Device Tools → Configuration → Process Measurement → Flow → Mass Flow Rate Unit
Field communicator Device Settings → Process Configuration → Mass Flow Rate → Mass Flow Rate Unit
Mass Flow Measurement Unit specifies the unit of measure that will be used for the mass flow rate. The
default unit used for mass total and mass inventory is derived from this unit.
Procedure
Set Mass Flow Measurement Unit to the unit you want to use.
Default: g/sec (grams per second)
Tip
If the measurement unit you want to use is not available, you can define a special measurement unit.
Options for Mass Flow Measurement Unit
The transmitter provides a standard set of measurement units for Mass Flow Measurement Unit, plus one
user-defined special measurement unit. Different communications tools may use different labels for the units.
Unit description
Label
Display ProLink III Field communicator
Grams per second gram/s g/sec g/s
Grams per minute gram/min g/min g/min
Grams per hour gram/h g/hr g/h
Kilograms per second kg/s kg/sec kg/s
Kilograms per minute kg/min kg/min kg/min
Kilograms per hour kg/h kg/hr kg/h
Kilograms per day kg/d kg/day kg/d
Metric tons per minute MetTon/min mTon/min MetTon/min
Metric tons per hour MetTon/h mTon/hr MetTon/h
Metric tons per day MetTon/d mTon/day MetTon/d
Pounds per second lb/s lbs/sec lb/s
Pounds per minute lb/min lbs/min lb/min
Configure process measurement Configuration and Use Manual
May2024 MS-00809-0200-1600
28 Emerson.com