13.1.4 Run a Smart Meter Verification test
Run a Smart Meter Verification Basic or Professional test to diagnose the flow meter (and flow meter system)
and verify if the flow meter is functioning properly and performing within factory specifications.
Important
Run the first Smart Meter Verification Basic or Professional test when the flow meter is installed in the pipeline
and the flow meter is at its normal operating conditions.
Run a Smart Meter Verification test using the display
Prerequisites
Read the Smart Meter Verification prerequisites in Use Smart Meter Verification.
Procedure
1. Choose Menu → Operations → Smart Meter Verification → Run Verification.
2. Select the desired output behavior.
Option
Description
Continue Measuring During the test, all outputs will continue to report their assigned process
variables. The test will run for approximately 90 seconds.
Fix at Last
Measured Value
During the test, all outputs will report the last measured value of their assigned
process variable. The test will run for approximately 140 seconds.
Fix at Fault During the test, all outputs will go to their configured fault action. The test will
run for approximately 140 seconds.
The test starts immediately.
3. Wait for the test to complete.
Note
At any time during the process, you can abort the test. If the outputs were fixed, they will return to
normal behavior.
Run a Smart Meter Verification test using ProLink III Basic or Professional
Procedure
1. Read the Smart Meter Verification prerequisites in Use Smart Meter Verification if you have not done so
already.
2. Run Smart Meter Verification Basic or Professional using ProLink III Basic or Professional:
• Smart Meter Verification Basic: Device Tools → Diagnostics → Meter Verification → Basic Meter
Verification
• Smart Meter Verification Professional: Smart Meter Verification Overview → Meter Verification →
Run Verification
• Smart Meter Verification Professional: Device Tools → Diagnostics → Meter Verification → Run Test
3. In the SMV Test Definition window, enter any desired information and click Next.
None of this information is required. It does not affect Smart Meter Verification processing.
ProLinkIII stores this information in the Smart Meter Verification database on the PC. It is not saved to
the transmitter.