11.4.1 Start, stop, and reset totalizers using the display
Prerequisites
To stop, start, or reset a single totalizer or inventory, the totalizer or inventory must be configured as a display
variable.
To reset an inventory using the display, this function must be enabled. To enable inventory reset using the
display, choose Menu → Configuration → Security → Display Security → Totalizers & Inventories →  Inventory
Reset and set Totalizer Reset to Allowed and save. Note that this affects only the display functions. Resetting
inventories using other tools is not affected.
Procedure
• To start or stop a single totalizer or inventory:
a) Wait or scroll until the totalizer or inventory appears on the display.
b) Choose Options.
c) Choose Start or Stop.
• To start or stop all totalizers and inventories as a group:
a) Choose Menu → Operations → Totalizers.
b) Choose Start or Stop.
• To reset a single totalizer or inventory:
a) Wait or scroll until the totalizer or inventory appears on the display.
b) Choose Options.
c) Choose Reset.
• To reset all totalizers and inventories as a group:
a) Choose Menu → Operations → Totalizers.
b) Choose Reset All.
11.4.2 Start, stop, and reset totalizers using ProLink III
Prerequisites
To reset an inventory using ProLink III, this function must be enabled. To enable inventory reset using ProLink
III, choose Tools → Options and enable Reset Inventories from ProLink III. Note that this affects only ProLink
III. Resetting inventories using other tools is not affected.
Procedure
• To start or stop a single totalizer:
a) Choose Device Tools → Totalizer Control → Totalizers.
b) Scroll to the totalizer that you want to start or stop, and click Start or Stop.
• To start or stop a single inventory:
a) Choose Device Tools → Totalizer Control → Inventories.
b) Scroll to the inventory that you want to start or stop, and click Start or Stop.
• To start or stop all totalizers as a group:
a) Choose Device Tools → Totalizer Control → Totalizers or Device Tools → Totalizer Control → 
Inventories.
b) Select Start All Totals or Stop All Totals.