4. Press OK to move through all test records in the local database.
View Smart Meter Verification test results from a Modbus host
Prerequisites
To read test results, the Modbus host must be connected to the transmitter.
Procedure
1. Specify the test record you want to read by writing a value between 0 and 19 to the appropriate
register.
Register Description
5779 Register 5825 = the most recent test
2. Read the Modbus register values using the descriptions in the following table.
Registers
Description
5697 Test number
5820 and 5821 Test result
0 = Pass
1 = Caution
5819 State
5818 Abort code (compressed)
5780 Time initiated
5782 and 5783 Inlet normalized data
5784 and 5785 Outlet normalized data
3. Disconnect as desired.
Interpreting Smart Meter Verification results
When the Smart Meter Verification Basic or Professional test is completed, the result is reported as Pass, Fail,
or Abort. (Some tools report the Fail result as Advisory instead.)
Pass
The meter is performing within factory specifications.
Abort
When you execute a Smart Meter Verification Basic or Professional test, the test performs a self-
diagnostic check to ensure that the flow meter is stable prior to running the test. In the rare case that
this check reveals an issue, Smart Meter Verification will report an abort code.
If you manually cancel an in-process Smart Meter Verification Basic or Professional test, the test
result displays Abort Code 1: User-Initiated Abort. In this case, you can restart Smart
Meter Verification without any further action. In the rare case any other abort occurs, contact factory
support.
In all cases where a Smart Meter Verification Professional test aborts, no report will be generated.
Fail
If a Smart Meter Verification Basic or Professional test ran at normal operations and failed, see
Resolve a failed Smart Meter Verification test to determine the appropriate actions.
Measurement support Configuration and Use Manual
May2024 MS-00809-0200-1600
142 Emerson.com