3. Set Flow Calibration Pressure to the pressure at which your sensor was calibrated.
The calibration pressure is the pressure at which your sensor was calibrated, and defines the pressure
at which there is no pressure effect. If the data is unavailable, enter 20 PSI.
4. Enter Flow Press Factor for your sensor.
The flow factor is the percent change in the flow rate per psi. When entering the value, reverse the sign.
Example
If the flow factor is −0.0002% per PSI, enter +0.0002 % per PSI.
5. Enter Density Pressure Factor for your sensor.
The density factor is the change in fluid density, in g/cm
3
/PSI. When entering the value, reverse the
sign.
Example
If the density factor is −0.000006 g/cm
3
/PSI, enter +0.000006 g/cm
3
/PSI.
6. Choose the method to be used to supply pressure data, and perform the required setup.
Method
Description Setup
Digital
communications
A host writes pressure data
to the meter at appropriate
intervals.
a. Using a field communicator, choose Configure
→ Manual Setup → Measurements → Optional Setup →
External Pressure/Temperature → Pressure.
b. Set Pressure Compensation to Enable.
c. Perform the necessary host programming and
communications setup to write pressure data to the
transmitter at appropriate intervals.
13.5 Validate the meter
Display
Menu → Configuration → Process Measurement → Flow Variables → Mass Flow Settings → Meter
Factor
Menu → Configuration → Process Measurement → Flow Variables → Volume Flow Settings →
Meter Factor
Menu → Configuration → Process Measurement → Density → Meter Factor
ProLink III Device Tools → Configuration → Process Measurement → Flow → Mass Flow Rate Meter Factor
Device Tools → Configuration → Process Measurement → Flow → Volume Flow Rate Meter Factor
Device Tools → Configuration → Process Measurement → Density → Density Meter Factor
Field communicator Device Settings → Process Configuration → Mass Flow → Mass Factor
Device Settings → Process Configuration → Volume Flow → Volume Factor
Device Settings → Process Configuration → Density → Density Factor
Meter validation compares flow meter measurements reported by the transmitter to an external
measurement standard. If the transmitter value for mass flow, volume flow, or density measurement is
significantly different from the external measurement standard, you may want to adjust the corresponding
meter factor. The flow meter’s actual measurement is multiplied by the meter factor, and the resulting value is
reported and used in further processing.
Changing a meter factor is very uncommon. Usually, it's done by official, legal proving companies. And the
meter factor is typically changed in the PLC, not at the meter.
Measurement support Configuration and Use Manual
May2024 MS-00809-0200-1600
152 Emerson.com