Option Description
Specify the hours until test
execution for a single execution
Write a floating-point value to Register 2993.
Specify the hours for a recurring
execution
• To specify the number of hours until the first test is executed,
write a floating-point value to Register 2993.
• To specify the number of hours between test executions,
write a floating-point value to Register 2995.
View the number of hours until the
next execution
Read Register 2997.
Cancel a scheduled execution • Write 0 to Register 2993.
• Write 0 to Register 2995.
2. Disconnect as desired.
Check for a field reference point
Use this procedure to check if a field reference point was created.
Prerequisites
• Smart Meter Verification Professional
• ProLink III Basic or Professional
Procedure
1. From ProLink III Basic or Professional, choose one of the following options:
• Device Tools → Diagnostics → Meter Verification → Run Test → View Previous Test Results
• Smart Meter Verification Overview → Meter Verification → History
2. Select Export Data to CSV File.
3. Save the CSV file to your computer.
4. Locate and open the CSV file.
5. Locate and examine the two columns labeled, Inlet Stiffness and Outlet Stiffness.
• If the field reference point has not been established, the numbers in both columns will be exactly 1.
• If the field reference point has been established, the numbers located in both columns will be close
to 1. The Inlet Stiffness and Outlet Stiffness numbers do not have to match.
13.2 Advanced Phase Measurement software
Micro Motion Advanced Phase Measurement software improves long-term flow reporting and measurement
performance in processes with intermittent periods of two-phase flow, including liquids with entrained gas
or gas with entrained liquid. If Advanced Phase Measurement is combined with the Net Oil or concentration
measurement software options, the software can also report liquid concentration, Net Oil, and/or Gas Void
Fraction (GVF) during the same two-phase conditions.
The following measurement options are available with Advanced Phase Measurement software:
• Net Oil
• Liquid with Gas
Configuration and Use Manual Measurement support
MS-00809-0200-1600 May2024