You cannot set these parameters from the display. If the default values are not appropriate, you must use
another method to set them.
Optional tasks in setting up concentration measurement
The following tasks are optional:
Modifying names and labels
Configuring extrapolation alerts
5.2.2 Set up concentration measurement using the display
This section guides you through most of the tasks related to setting up and implementing the concentration
measurement application.
Restriction
This section does not cover building a concentration matrix. For detailed information on building a matrix, see
the Micro Motion Enhanced Density Application Manual.
Enable concentration measurement using the display
The concentration measurement application must be enabled before you can perform any setup. If the
concentration measurement application was enabled at the factory, you do not need to enable it now.
Prerequisites
The concentration measurement application must be licensed on your transmitter.
Disable the following applications before enabling concentration measurement as concentration
measurement cannot be enabled at the same time:
Advanced Phase Measurement — gas with liquid
API Referral
Gas Standard Volume
Procedure
1. Choose MenuConfigurationProcess Measurement.
2. Choose Flow VariablesVolume Flow Settings and ensure that Flow Type is set to Liquid.
3. Return to the Process Measurement menu.
4. If the API Referral application is displayed in the menu, choose API Referral and ensure that Enabled/
Disabled is set to Disabled.
The concentration measurement application and the API Referral application cannot be enabled
simultaneously.
5. If the Advanced Phase Measurement application is displayed in the menu, choose Advanced Phase
MeasurementApplication Setup and ensure that Enabled/Disabled is set to Disabled.
6. Enable concentration measurement.
a) Choose MenuConfigurationProcess MeasurementConcentration Measurement.
b) Set Enabled/Disabled to Enabled.
Set up temperature data using the display
The concentration measurement application uses line temperature data in its calculations. You must decide
how to provide this data, then perform the required configuration and setup. Temperature data from the
Configuration and Use Manual Configure process measurement applications
MS-00809-0200-1600 May2024
Configuration and Use Manual 61